Protocole for evaluation and use of a low-cost particulate matter monitoring station
Protocole d'évaluation et d'utilisation d'une station low-cost de mesure des particules fines
Résumé
The use of low-cost sensors could revolutionize the measurement of fine particles (PM). Therefore, we must evaluate their performance. Here we report the assembly, evaluation and use of a station using a PMS7003 sensor. Its performance to measure PM2.5 (PM inferior to 2.5µm) is compared to the recommendations of the US Environmental Protection Agency (EPA). Some metrics (slope, NRMSE) are not compliant. A linear calibration brings them within EPA limits and a mechanistic calibration further improves accuracy. We can achieve RMSEs of 0.90 µg/m3 (hourly scale) and 0.48 µg/m3 (daily scale).
Les capteurs « low-cost » pourraient révolutionner la mesure des particules fines (PM) mais il paraît nécessaire d'évaluer leurs performances. Ce travail propose un protocole de montage, d'évaluation et d'utilisation d'une station comportant un capteur PMS7003. Les performances de mesure des PM2.5 (PM de taille <2,5µm) sont comparées aux directives de l'Environmental Protection Agency américaine (EPA). Il apparaît que certaines métriques (pente et NRMSE) sont non conformes en l'absence de calibration. Une calibration linéaire les ramène dans les limites de l'EPA et une calibration mécanistique améliore encore la précision. Les RMSEs sont alors de 0.90 µg/m 3 (échelle horaire) et 0.48 µg/m 3 (journée).
Origine | Fichiers produits par l'(les) auteur(s) |
---|